Fiche publication


Date publication

mai 2018

Journal

Physics in medicine and biology

Auteurs

Membres identifiés du Cancéropôle Est :
Pr BRUNOTTE François


Tous les auteurs :
Vrigneaud JM, McGrath J, Courteau A, Pegg R, Sanchez-Pastor Gomis A, Camacho A, Martin G, Schramm N, Brunotte F

Résumé

We evaluated the performance characteristics of a prototype preclinical PET scanner available as an easy clippable assembly that can dock to an MRI system. The single ring version of the PET system consists of 8 detectors, each of which comprises a 12 × 12 silicon photomultipliers (SiPMs) array coupled with a dual layer of offset scintillation crystals to measure depth of interaction. The crystal arrays have 29 × 29 (30 × 30 for the outer layer) 4 mm long LYSO crystals (6 mm for the outer layer). The ring diameter is 119.2 mm and the axial field of view is 50.4 mm. The NEMA NU-4-2008 protocol was followed for studying the PET performance. Temperature stability of SiPMs was also investigated. The peak system absolute sensitivity was 4.70% with an energy window of 250-750 keV. The spatial resolution was 1.28/1.88/1.85 mm FWHM (radial/tangential/axial) at a distance of 5 mm from the center. Peak noise equivalent counting rate (NECR) and scatter fraction for mouse phantom were 61.9 kcps at 14.9 MBq and 21.0%, respectively. The uniformity was 6.3% and the spill-over ratios in the images of the water- and air-filled chambers were 0.07 and 0.17, respectively. Recovery coefficients ranged from 0.13 to 0.96. Change in sensitivity as a function of ambient temperature was 0.3%/°C. These first results indicate excellent spatial resolution performance for use with animal studies. Moreover, the clippable assembly can be upgraded to accept a second ring of SiPMs modules, leading to improved sensitivity and axial coverage.

Référence

Phys Med Biol. 2018 May 15;: