Fiche publication
Date publication
septembre 2020
Journal
Lab on a chip
Auteurs
Membres identifiés du Cancéropôle Est :
Dr BOLOPION Aude
,
Dr GAUTHIER Michaël
Tous les auteurs :
Daguerre H, Solsona M, Cottet J, Gauthier M, Renaud P, Bolopion A
Lien Pubmed
Résumé
Microfluidic electrical impedance flow cytometry is now a well-known and established method for single-cell analysis. Given the richness of the information provided by impedance measurements, this non-invasive and label-free approach can be used in a wide field of applications ranging from simple cell counting to disease diagnostics. One of its major limitations is the variation of the impedance signal with the position of the cell in the sensing area. Indeed, identical particles traveling along different trajectories do not result in the same data. The positional dependence can be considered as a challenge for the accuracy of microfluidic impedance cytometers. On the other hand, it has recently been regarded by several groups as an opportunity to estimate the position of particles in the microchannel and thus take a further step in the logic of integrating sensors in so-called "Lab-on-a-chip" devices. This review provides a comprehensive overview of the physical grounds of the positional dependence of impedance measurements. Then, both the developed strategies to reduce position influence in impedance-based assays and the recent reported technologies exploiting that dependence for the integration of position detection in microfluidic devices are reviewed.
Référence
Lab Chip. 2020 Sep 11;: