Fiche publication
Date publication
mars 2021
Journal
Medical physics
Auteurs
Membres identifiés du Cancéropôle Est :
Dr QUETIN Philippe
,
Dr RETIF Paul
Tous les auteurs :
Retif P, Djibo Sidikou A, Letellier R, Verrecchia-Ramos E, Quetin P
Lien Pubmed
Résumé
A common dosimetric quality-assurance (QA) method in stereotactic body radiation therapy (SBRT) of lung tumors is to use lung phantoms with radiochromic film. However, in most phantoms, the film moves with the tumor, leading to the blurring effect. This technical note presents the QA performance of a novel phantom in which the film is fixed; this phantom can be used for both patient-specific QA and end-to-end testing.
Mots clés
3D-printed, film dosimetry, lung SBRT, tumor motion
Référence
Med Phys. 2021 Mar 15;: