Fiche publication


Date publication

mars 2021

Journal

Medical physics

Auteurs

Membres identifiés du Cancéropôle Est :
Dr QUETIN Philippe , Dr RETIF Paul


Tous les auteurs :
Retif P, Djibo Sidikou A, Letellier R, Verrecchia-Ramos E, Quetin P

Résumé

A common dosimetric quality-assurance (QA) method in stereotactic body radiation therapy (SBRT) of lung tumors is to use lung phantoms with radiochromic film. However, in most phantoms, the film moves with the tumor, leading to the blurring effect. This technical note presents the QA performance of a novel phantom in which the film is fixed; this phantom can be used for both patient-specific QA and end-to-end testing.

Mots clés

3D-printed, film dosimetry, lung SBRT, tumor motion

Référence

Med Phys. 2021 Mar 15;: