Fiche publication
Date publication
avril 2023
Journal
Sensors (Basel, Switzerland)
Auteurs
Membres identifiés du Cancéropôle Est :
Dr WACOGNE Bruno
Tous les auteurs :
Wacogne B, Belinger Podevin M, Vaccari N, Koubevi C, Codjiová C, Gutierrez E, Bourgeois P, Davoine L, Robert-Nicoud M, Rouleau A, Frelet-Barrand A
Lien Pubmed
Résumé
Despite numerous innovations, measuring bacteria concentrations on a routine basis is still time consuming and ensuring accurate measurements requires careful handling. Furthermore, it often requires sampling small volumes of bacteria suspensions which might be poorly representative of the real bacteria concentration. In this paper, we propose a spectroscopy measurement method based on a description of the absorption/attenuation spectra of ESKAPEE bacteria. Concentrations were measured with accuracies less than 2%. In addition, mixing the mathematical description of the absorption/attenuation spectra of mammalian T-cells and bacteria allows for the simultaneous measurements of both species' concentrations. This method allows real-time, sampling-free and seeder-free measurement and can be easily integrated into a closed-system environment.
Mots clés
ESKAPEE bacteria, Lactococcus lactis, absorption spectra, co-culture monitoring, contamination, mammalian T-cells, white light spectroscopy
Référence
Sensors (Basel). 2023 04 27;23(9):