Fiche publication
Date publication
décembre 2011
Auteurs
Membres identifiés du Cancéropôle Est :
Pr MARESCAUX Jacques
,
Pr SOLER Luc
Tous les auteurs :
Diana M, Wall J, Perretta S, Dallemagne B, Gonzales KD, Harrison MR, Agnus V, Soler L, Nicolau S, Marescaux J
Lien Pubmed
Résumé
OBJECTIVE: This study aimed to assess the feasibility of a totally endoscopic enteral bypass using a self-orienting, dual ring, magnetic anastomosis system (MAGNAMOSIS) guided by a magnetic tracking system (3D METRIS). MATERIALS AND METHODS: In an anesthetized pig, 2 endoscopes were advanced, one each into the stomach and the colon. Both endoscopes were equipped with a MAGNAMOSIS ring secured with an endoscopic snare and a 3D METRIS within one working channel. The whole procedure was followed laparoscopically. The tracking system guided tips of endoscopes to a "rendez-vous" location between the colon and stomach. RESULTS: MAGNAMOSIS magnets automatically joined in the correct configuration when guided to within 2 cm of each other. At necropsy, magnetic rings were secure without entrapment of excess bowel or mesentery. CONCLUSION: An endoscopic enteral bypass with magnetic anastomosis and magnetic tracking device was feasible. More accurate tracking and advanced techniques could enable endoscopic bypasses at multiple sites in the gastrointestinal tract.
Référence
Surg Innov. 2011 Dec;18(4):317-20