Quantitative analysis by high-performance liquid chromatography atmospheric pressure chemical ionization mass spectrometry: The determination of the renin inhibitor CP-80,794 in human serum.
Fiche publication
Date publication
avril 1991
Auteurs
Membres identifiés du Cancéropôle Est :
Pr SCHNEIDER Raphaël
Tous les auteurs :
Fouda H, Nocerini M, Schneider R, Gedutis C
Lien Pubmed
Résumé
A specific method was developed for the quantitative determination of the renin inhibitor CP-BO,794. Serum extracts containing the drug and an internal standard were injected into a standard reverse-phase high-performance liquid chromatography (HPLC) column. The mobile phase, methanol/water (8/2), flowed at 1 ml/min through the column and then via a heated nebulizer interface into a corona discharge atmospheric pressure chemical ionization source. The assay minimum limit of quantification was 50 pg/mL. It exhibits satisfactory accuracy and precision over the range 50 pg/ml, to 10 ng/mL. A minor modification of the HPLC mobile phase was necessary to attain extremely low detection limits. The addition of a structural analogue contributed to enhancing the precision of the assay.
Référence
J Am Soc Mass Spectrom. 1991 Apr;2(2):164-7